rAPID-T Discrete Analyzer

The rAPID-T software/hardware platform offers a compact, affordable automated Chemistry Analyzer.

  • Durable. Accurate. Precise. Economical
  • Open software platform. Program it to run:
  • Up to 100+ tests per hour with 10mm path length cuvettes
  • Built-in temperature controlled reactions
  • Wavelengths: 340, 405, 420, 480, 505, 540, 570, 660 or … custom
  • Small footprint and weight: 21” W x 15.5” H x 20” D and 37 pounds!
  • Compatible with Windows Vista or 7/8/10 (computer not supplied)

The software optimizes the rAPID-T's advanced hardware and performance. Operators can program their colorimetric applications into rAPID-T’s easy-to-use method setup and configurations. Because of rAPID-T’s open platform, operators can also adapt their system to use assay kits for Water, Wine, Food and other analyses from several manufacturers (e.g. Megazyme, Bio-Rad, Thermo Scientific, NZYTech, Nitrate Elimination Company, etc.) 

Specifications

Typical throughput up to 100+ tests per hour
Typical reaction volume 250 µL
Path length of cuvette 10mm
Reaction vessel 5-well cuvettes, disposable
Maximum reaction volume 500 µL
Dimensions 21” (53cm) W x 15.5” (40cm) H x 20” (50 cm) D
Weight 37 lbs (17 kg)

Reagent and Sample Handling

Syringe 500 μL, range 2 μL – 450 μL w/increments of 0.5 μL
Probe 316 stainless steel w/liquid level sensing
Mixing Probe oscillation, time and speed adjustable
Precision for volumes < 5 μL < 3% CV
Precision for volumes > 5 μL < 2% CV
No. of reaction cuvette places 40 w/PAUSE feature to allow operator to add more during a run
No. of reagent places 10 Total: 15 mL and 30 mL
No. of sample places 40 places for 4 mL sampler cups w/PAUSE feature to add additional samples during a run
Reaction temperature From ambient to 40⁰C
  Incubation is software controlled and automatically optimized

Reading

Optical design  User-selected monochromatic or bichromatic results
  8-position filter wheel
  Wavelengths: 340, 405, 420, 480, 505, 540, 570, 660 or custom
Interference filters  Long life, hard coat, ion assisted deposition, ~10nm half bypass
Linear range  0.0 to 3.0 A
Photometer accuracy + (1% reading + 0.005A) from 0 to 1.0 A
  + (2% reading + 0.005A) from 1.0 A to 3.0 A

Software

Format  USB with internet upgrades
QC Functions  Store Control Data, QC Data Charting Screen
Operating system  Windows® 7, 8, 8.1 or 10 Professional
System Diagnostic Tools  Probe/Rack Alignment, Control Scripts, Self-Test, QC Well
Self-monitoring modes  Lamp, filters, reagent levels, Cuvette

Certifications

NRTL listed (Canada and US), CE marked

Manufactured by ISO 13485:2003 certified manufacturer in the USA.